Peak problems



LevelBasic

Symptoms:

  • No peaks
  • Very small peaks
  • Back tailing peaks
  • Front tailing peaks
  • Split / double peaks
  • Round peaks
  • Negative peaks
  • Ghost peaks

 

No peaks

NO PEAKS

Possible causes Solution
Detector lamp is defect. Replace the lamp.
Detector or data system settings fault.


Apply correct values.
Compounds too strongly retained. Try a higher eluent strength.

Very small peaks.

Very small peaks
Possible causes Solution
Sample deterioration Prepare fresh samples or check stockage conditions
Detector or recorder settings too high. Check all values in the data system.
Pollution of the detector cell. Rinse the detector cell with clean solvents.

Back tailing peaks.

Back tailing peaks
Possible causes Solution
Extra column back broadening.

Make sure the tubing length and tubing id are kept to a minimum.
Keep in mind that the column outlet tubing has more back broadening influence than the column inlet tubing.


Checked detector flow cell volume: must be smaller than 0.1 x peak volume.
Check the influence of the detector's time constant: keep this at < 0.5 sec.

Mobile phase pH not correct or optimal. Check and change
Column overloading. Try injecting less sample or a lower concentration.
Basic analytes with active residual silanols. Use another stationary phase with higher surface coverage or try adding triethylamine to the eluent (usually < 1 ml/l).

Front tailing peaks.

Front tailing peaks
Possible causes Solution
Column overloaded. Try injecting less sample or a lower concentration.
Sample solvent not compatible with eluent. Try to dissolve the sample in mobile phase.

Split peaks.

Split peaks
Possible causes Solution
Contamination of guard or analytical column. Replace the guard column. Regenerate the analytical column.
Sample solvent not compatible with eluent. Try to dissolve the sample in mobile phase.

Round (double) peaks.

Round (double) peaks
Possible causes Solution
Sample concentration higher than linear detector range. Dilute the sample.
Column overloaded.  Dilute the sample.
Co-elution of 2 components. Improve separation. Change conditions?
Poor column bed stability Backflush or replace the column.

Negative peaks.

Negative peaks
Possible causes: Solution
Refractive index of analytes less than that of the eluent (with RI detection). Flush reference detector cell with mobile phase.

Eluent absorbs more uv than the analytes (indirect uv detection).

Try another UV wavelength.

Ghost peaks.

Ghost peaks
Possible causes Solution
Contamination in injector from previous sample (cross contamination). Flush more with stronger solvents between injections. Reduce sample concentration, if possible.
When continuous degassing with helium is applied: contaminants from the helium coming our of the helium itself or out of the gas tubing.


 

Use a gas clean filter in the helium line.

Click here to go to the Home-page