Baseline problems



LevelBasic

Baseline noise.

Baseline noise
Possible causes Solutions
Electrical problem. Take care that the pump, detector and data system are all set at the correct values and are functioning well.
Bubbles in detector cell.

First degas the mobile phase solvents properly and check your degassing system.

 

Connect detector directly to the pump and flush out the bubble. It is also possible to compress the bubble by increasing the pressure at the detector outlet with a restrictor or simply with a finger. Be careful and do not apply a high pressure when the flow cell is not strong enough.

Detector lamp old. Replace the lamp.
Eluent not sufficiently degassed. Continually degas the eluent.
Spikes caused by bubbles coming forth out of heat development by mixing two solvents, like methanol and water. Check / improve solvent degassing.

Unstable baseline.

Unstable baseline
Possible causes Solutions
Unstable gradient or mixing. Check gradient pump, degassing, propotional valves and mixing chamber.
Temperature fluctuations. Use column thermostat. Column temperature > 5 ° above ambient.
Bubbles in the detector cell Degas mobile phase properly. Apply back pressure behind detector cell (check manual if possible!) to avoid bubble formation behind the column.
Poor gradient mixing. Run an blank gradient with methanol (+ 0.1 % acetone ) to locate the problem.

Baseline drift.

Baseline drift
Possible causes Solutions
Eluent impurities Replace solvents. Use HPLC grade only.
Temperature fluctuations. Use or check the column oven. Adjust the oven temperature at least 5 degrees above room temperature.
A change in the equilibrium inside the column (impurities). Regenerate column and use fresh eluent.
Gradient elution. Detection level of eluent changes with change in composition.

Click here to go to the Home-page